Instruction Manual PCD 650
46
3.3 pH Calibration
3.3.1 pH Calibration with a Standard Buffer
Make sure you have selected a standard buffer with which you intend to perform pH
calibration. Refer ’pH Setup’ section in page 82 for more details of setting buffer types.
To start pH Calibration:
1. Switch on the meter and make sure the meter is in pH measurement mode.
2. Press CAL (F2) to start calibration.
3. The meter shows ‘Calibration-Rinse Electrode’ screen [Figure 41 (a)] for
few seconds and then shows the pH calibration screen [Figure 41-(b)].
(a) (b)
Figure 41: pH Calibration Screens
4. Rinse the electrode in clean water and prepare electrode for calibration.
Refer ‘Prepare the meter for measurement ‘section in page 40.
Note: If the meter is password protected, you will be prompted to enter a
password. Refer ‘Accessing Calibration mode when password protection
enabled’ in pa
g
e 42.
Note: If you wish to completely re-calibrate the meter, you need to clear
previous calibration data. Press CLR-C (F2) key to clear previous
calibration. The meter shows you confirmation screen. Press ENTER key
to confirm deleting previous calibration. Once cleared, you need to
calibrate the first point that corresponds to your selected buffer group, as
mentioned on page 39.