Instruction Manual PCD 650
17
2.2 Taking Measurement
2.2.1 Prepare the meter for measurement
Before you start measuring,
Make sure you have connected a suitable probe (pH/OPR/ISE),
Conductivity probe and DO with in-built ATC to the meter.
Make sure the probes are in good working condition & clean. If required,
clean (pH/OPR/ISE) probe with clean water, conductivity and DO probe
with de-ionized water to remove impurities. For more information on
probe’s conditioning refer page 7.
Make sure batteries have been installed or the DC adapter is connected to
the meter.
Perform calibration if you change to a new probe.
2.2.2 Taking a reading
1. Press ON (F4) key to switch on the meter. The ON (F4) key has to be kept
pressed until the display comes up.
2. Make sure you are in the required measurement mode. Press MODE (F3) to
switch between modes.
3. Dip the appropriate probe with built in ATC probe into the sample solution.
4. The LCD shows ‘Stable’ indicator if this feature is enabled in setup. (See
page 75).
5. Note the reading.
2.2.3 Stable reading indicator
You can configure the meter so that LCD displays a ‘Stable’ indicator when
the reading does not vary for 2 consecutive seconds. The amount of
variations allowed can be set as ‘Slow’, ‘Medium’ or ‘Fast’. (See page 75)
2.2.4 Holding a reading
In some situations, you may want to freeze (hold) the measured reading in
the LCD for a delayed observation. You can hold a reading in two different
ways.
Manual Hold – Allows you to hold the reading by pressing HOLD (F1) key at
any time you want. When you hold a reading, the ‘HOLD’ indicator starts
flashing. The readings (including temperature reading) will be held until you
press any other key again.
Auto-Hold –The meter automatically holds the reading if the reading is
Stable’ for 5 consecutive seconds. This feature needs to be enabled in the
setup (See page 75). Press HOLD (F1) key to release the reading.
Note: When dipping the probe into sample, the sensor or the glass bulb of the
probe must be completely immersed into the sample. Stir the probe gently in the
sample to create a homogeneous sample. Allow time for the reading to stabilize.