EUTECH INSTRUMENTS CYBERSCAN DO 5000 DO METER Bedienerhandbuch Seite 45

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 76
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 44
Instruction Manual PD 300
8.7 P2.0 (OFS): Offset for % Saturation
Measurement
Figure 30: Offset for % saturation
NOTE
This sub group appears in % saturation measurement
mode only.
From measurement mode:
1. Press the MODE key to select % saturation mode.
2. Press the SETUP key to enter SETUP mode.
3. Press or keys to scroll through su/jointfilesconvert/267593/bgroups
until you view parameter “OFS” in the upper
display.
4. Press the ENTER key. The upper display shows
the current measurement in % saturation and the
lower shows “OFS”.
5. Press or keys to offset the % saturation
measurement.
6. Press the ENTER key to confirm selection and to
move back to su/jointfilesconvert/267593/bgroup “OFS”. If you want to
return to measurement mode, press CAL/MEAS.
(The values shown above are for illustration purpose. Your meter may
display different values depending on the settings & your environmental
conditions)
40
Seitenansicht 44
1 2 ... 40 41 42 43 44 45 46 47 48 49 50 ... 75 76

Kommentare zu diesen Handbüchern

Keine Kommentare