EUTECH INSTRUMENTS ION 700 - REV 3 Bedienerhandbuch

Stöbern Sie online oder laden Sie Bedienerhandbuch nach Messen, Testen & Steuern EUTECH INSTRUMENTS ION 700 - REV 3 herunter. EUTECH INSTRUMENTS ION 700 - REV 3 Instruction manual Benutzerhandbuch

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 36
  • Inhaltsverzeichnis
  • FEHLERBEHEBUNG
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 0
Instruction Manual
DO 700 Dissolved Oxygen/ºC/ºF
Bench Meter
Technolo
g
y
M
ade
E
a
s
y
.
.
.
Part of Thermo Fisher Scientific 68X541703 Rev. 1 Feb. 2010
Seitenansicht 0
1 2 3 4 5 6 ... 35 36

Inhaltsverzeichnis

Seite 1 - Instruction Manual

Instruction Manual DO 700 Dissolved Oxygen/ºC/ºF Bench Meter TechnologyMadeEasy... Part of Thermo Fisher Scientific

Seite 2

3. DISSOLVED OXYGEN / ATC ELECTRODE The DO 700 includes a galvanic DO electrode with built in temperature measurement. The electrode has a dual cab

Seite 3

5 4. CALIBRATION The DO 700 has three measurement modes; DO as % saturation, DO as mg/L (or ppm) concentration, and temperature. Dissolved oxygen l

Seite 4 - TABLE OF CONTENTS

The DO 700 will accept one calibration point in concentration mode. The minimum value is 2 mg/L (ppm), and the calibration window is +/- 40% of the fa

Seite 5

4.2 % Saturation Calibration The DO700 can be easily calibrated in air. For best accuracy, ensure that the barometric pressure value is accurate. Th

Seite 6

4) Press . The CAL indicator will blink and the secondary display will indicate the calibration point, 0.0. 5) Press to confirm the calibration. Th

Seite 7

9 NOTE: The minimum calibration value is 2 mg/L. In addition, the calibration is limited to ±40% adjustment of the factory default value to prevent

Seite 8 - 2. GETTING STARTED

5. SETUP FUNCTIONS Use the setup feature to customize your instrument operation. There are two setup menus; one menu is derived from the % saturation

Seite 9

Press to view Offset % (OFS)—% saturation mode only. 5.3 3.0 ConF (Configuration) Press to access set-up programs 3.1 thru 3.5. 5.4 3.1 rdY (Re

Seite 10

5.7 3.4 SALt (Salinity Adjustment) For best accuracy, enter the actual salinity value if your samples are at least 0.1 ppt (100 ppm). Maximum adjust

Seite 11 - 4. CALIBRATION

5.10 4.0 rSt (Reset) Press . Press or to select “Yes” (Reset) or “No” (Cancel). If “Yes”, press or to select “Cal” (calibration reset only) or “

Seite 13

6. TAKING MEASUREMENTS During measurement the dissolved oxygen reading is automatically compensated for pressure and salinity. For best accuracy, ens

Seite 14

6.1 HOLD Function For prolonged observation of a reading, press during measurement mode to freeze the display. The “HOLD” indicator will display whe

Seite 15

16 7. ELECTRODE MAINTENANCE The DO 700 electrode produces an output proportional to the oxygen present in the medium in which it is placed. The gal

Seite 16 - 5. SETUP FUNCTIONS

There are no special probe storage requirements. Cap and Electrolyte Replacement Replacement of the pre-assembled cap is required only when you canno

Seite 17 - % saturation mode only

18 Electrolyte Solution The electrolyte solution in your probe’s cap will deplete on usage and will need to be replaced periodically. The replacemen

Seite 18 - Press to confirm

8. DISSOLVED OXYGEN THEORY Dissolved Oxygen (DO) refers to the volume of oxygen that is contained in water. There are two main sources of DO in wate

Seite 19

20 Measurement Units One measure of DO in water is parts per million (ppm) which is the number of oxygen molecules (O2) per million total molecules

Seite 20

21 Air Calibration Understanding the principle of air calibration is easy, once you know that it is partial pressure that the probe is responding t

Seite 21

22 Some pollutants such as acid mine drainage produce direct chemical demands on oxygen in the water. DO is consumed in the oxidation-reduction rea

Seite 22 - 7. ELECTRODE MAINTENANCE

23 9. REPLACEMENTS AND ACCESSORIES Part number Ordering Code Item Description Eutech Instruments Oakton Instruments DO 700 meter with DO/ATC probe

Seite 24 - Electrolyte Solution

10. TROUBLESHOOTING GUIDE PROBLEM CAUSE SOLUTION No display Main power not switched on. AC Adapter socket not inserted properly. Switch on the powe

Seite 25 - Temperature °C

PROBLEM CAUSE SOLUTION CAL % calibration value is not acceptable. Electrode is disconnected or failing. Check electrode connections. Check that c

Seite 26

26 11. SPECIFICATIONS DO % Saturation Range Resolution Relative accuracy Offset adjustment Calibration Points 0 to 300% 0.1 % (0.0 to 199.9%);

Seite 27

27 12. WARRANTY This meter is supplied with a warranty against significant deviations in material and workmanship for a period of THREE years from

Seite 28

28 13. RETURN OF ITEMS Authorization must be obtained from our Customer Service Department or authorized distributor before returning items for any

Seite 30 - 10. TROUBLESHOOTING GUIDE

For more information on our products, please contact our channel partner or visit our websites listed below: Oakton Instruments 625 E Bunker

Seite 31

TABLE OF CONTENTS 1. INTRODUCTION ... 1 2. GETTING STARTED...

Seite 32 - Salinity Correction

9. REPLACEMENTS AND ACCESSORIES... 23 10. TROUBLESHOOTING GUIDE... 24 11. SPECIFICATION

Seite 34 - 13. RETURN OF ITEMS

1 1. INTRODUCTION Thank you for purchasing our DO 700 series benchtop meter. This microprocessor-based meter is economical and simple to use. The d

Seite 35

2. GETTING STARTED 2.1 Keypad Functions Powers the meter on and off. Upon power on, the meter automatically begins in the mode that was last used.

Seite 36

2.2 LCD Annunciators 2.3 Meter Connections DC Power supply ATC Temperature connection of Dissolved Oxygen electrode DO BNC connection of Disso

Kommentare zu diesen Handbüchern

Keine Kommentare