Instruction Manual DO 300
24
8.4 OFS: Offset for % Saturation Measurement
NOTE: This feature appears from % saturation SETUP mode only.
From measurement mode:
1. Press the MODE key to select % saturation mode.
2. Press the SETUP key to enter SETUP mode.
3. Press MI/ or MR/ keys to scroll through main groups
until you view parameter “OFS” in the upper display.
4. Press the ENTER key. The upper display shows the
current measurement in % saturation and the lower
shows “OFS”.
5. Press MI/ or MR/ keys to offset the % saturation
measurement.
6. Press the ENTER key to confirm selection and to move
back to main group “OFS”. If you want to return to
measurement mode, press CAL/MEAS.
(The values shown above are for illustration purpose. Your meter may display
different values depending on the settings & your environmental conditions)
Figure 20: Offset for %
saturation