Instruction Manual alpha-pH 1000
19
7.2.2 Temperature calibration (ATC mode only)
1) Select “ATC on” as described above in Section 7.2.1.
2) Press the ENTER key. The upper display indicates the current
temperature offset. The current measured temperature is shown in
the lower display.
3) Compare the current measured temperature on the controller
display to a thermometer known to be accurate. Note down the
correct temperature value.
4) Press the ▲ or ▼ key to scroll the lower display to match the
correct value. The upper display will now show the offset value. You
can offset temperature up to +/- 5
o
C.
5) Press the ENTER key to confirm your selection.
6) Continue with additional Advanced Set-up procedures, or return to
the Measurement mode by pressing the ▲ and ▼ keys (escape)
simultaneously.
7.2.3 Setting manual temperature compensation
Note: This parameter is blanked out when the controller is set for ATC
operation.
For manual temperature compensation, you can set two different
temperatures: process and calibration. This allows calibration at a
temperature other than your process temperature. Example: setting a
calibration temperature of 25
o
C lets you calibrate using standard buffer
solutions at 25
o
C, even if your process temperature is a different
temperature.
1) Select “ATC oFF” as described above in section 7.2.1
2) Press the ENTER key. The upper display shows the current
process temperature and the lower display shows “P.
o
C” to
indicate process temperature.
3) Press the ▲ or ▼ key to adjust the process temperature value.
You can adjust the value from –9.9 to 125
o
C.
4) After you set the process temperature value, press the ENTER
key. The upper display shows the current calibration temperature
and the lower display shows “C.
o
C” to indicate calibration
temperature.
5) Press the ▲ or ▼ key to adjust the calibration temperature value. You can adjust this value from –9.9 to
125
o
C.
6) Press the ENTER key to confirm your selection.
7) Continue with additional Advanced Set-up procedures, or return to Measurement mode by pressing the
▲ and ▼ keys (escape) simultaneously.
0.0
25.0
ATC
SETUP
HOLD
o
C
on
tc
SETUP
HOLD
25.0
C.C
SETUP
HOLD
oFF
tc
SETUP
HOLD
25.0
P.C
SETUP
HOLD