Instruction Manual alpha-pH 1000
16
Note: If there is a calibration error, the controller displays “ERR”. If this happens, push both the ▲ and ▼
keys (escape) to restart the calibration beginning from step 1.
Note: When calibrating with manual temperature compensation, the controller automatically changes from
the preset process temperature to the calibration temperature. After leaving the Calibration mode, the
controller switches back to process temperature (for setting the calibration temperature and the process
temperature, see section 7.2.3).
6.2 ORP – mV Calibration
This mode allows one-point calibration.
1) Enter Calibration mode. While in Measurement mode, push
the CAL key and scroll to Calibration Code “11”. Push the
ENTER key. The upper and lower display reads “CAL OrP”.
Note: If the upper and lower display read “CAL PH”, see section
7.6.2 for procedures on how to switch from PH to ORP mV
readings.
2) Place sensor in the ORP solution.
3) Press the ENTER key to begin calibration. The “CAL”
indicator appears on the display. The upper display shows the
current mV output of the electrode without any offset
adjustment.
4) Press the ▲ or ▼ key to match the mV value to your exact
reading.
5) To confirm, press the ENTER key. The controller calculates
the correction factor and returns to the Measurement mode.
CAL
OrP
SETUP
HOLD
226
mV
C
HOLD
CAL