EUTECH INSTRUMENTS alpha-pH 2000D Bedienerhandbuch Seite 49

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 52
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 48
Instruction Manual αlpha-pH2000D
- 43 -
11.4.5 Control signal of Pulse Frequency Controllers
Relay
ON
tON
tOFF
Pulse Length T
OFF
Time [s]
The output relay of the pulse frequency controller is clock-timed. The pulse duration tON
remains constant at 250 msec. Depending on the divergence from the limit value, the
frequency (1/T) is increased or decreased in accordance with the proportional range Xp.
The following applies:
tON = Const. (250 msec.)
greater divergence Î greater f (greater frequency)
Xp exceeded Î max. frequency
Seitenansicht 48
1 2 ... 44 45 46 47 48 49 50 51 52

Kommentare zu diesen Handbüchern

Keine Kommentare