WATER QUALITY
INSTRUMENTS
18
EnviroEquip Flow Cell
The EFC500 ow cell is a simple open cup type cell that
hangs o the inside of a bucket or other container. The
water quality probes are simply placed loosely into the cell.
A lid is also provided with holes set up to hold sensors from
the TPS 90FLMV water quality instrument.
Smaller volume - faster sampling
•
Easy to use - no extra ttings or joiners •
Counter-ow - no sample reaction with air •
Easy to set up - just hang in a bucket •
Easy to clean - comes apart with no tools •
3 way valve for accurate ow collection •
Stainless steel ttings•
When the pump is turned on, water is pumped through the
cell and overows the top. This ensures all the probes are
fully submersed in water. Also since the ow cell is hanging
inside a bucket, you don’t need to worry about controlling
the out-ow from the cell to avoid spills.
When pumping is rst started, the three 3 valve is set to
ow through the valve and into the ow cell and NOT out
the sample tube on the branch leg of the 3 way valve. As
water is pumped in through the inlet port, it lls the ow
cell and then it drains out over the top into the bucket.
When your water quality parameters have stabilised to
meet your criteria, then redivert the 3 way valve and let the
sample water ow directly out the branch leg of the 3 way
valve and into your sample collection vessel. This bypasses
the ow cell and eliminates any time averaging that would
be occurring in the cell volume.
The ow cell has a volume of 500 ml empty and approx 400
ml with the probes inserted. This small volume will reduce
the pumping time to get fresh formation water pumped
through the cell.
Ordering Information
EEKEFC500EnviroEquip Flow Cell, 500ml with lid