© 05/07 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Transmitter Cond/Res 2-Channel M300
Printed in Switzerland 52 121 318
7 Sensor Calibration
(PATH: Cal)
The calibration key allows the user one-touch access to sensor calibration and verification
features. When performing calibration on a conductivity or resistivity sensor, results will vary
depending on the methods, calibration apparatus and/or quality of reference standards used to
perform the calibration.
7.1 Enter Calibration Mode
While in Measurement mode press the key.
Press the or key to select the type of calibration desired. The options are ‘Meter’, ‘Analog’,
and ‘Sensor’.
C
NOTE: Dual channel transmitters allow ONLY the calibrate sensor option in the default
mode. For access to all calibration functions see section 11.2.3.
After selecting ‘Sensor’, press Enter and the cursor moves to the next line to select the desired
Sensor Calibration task, The choices are ‘Conductivity’, ‘Resistivity’, ‘Temperature’, ‘Edit’ and
‘Verify’. A flashing ‘H’ in the upper left corner of the display indicates a calibration is in process
with a Hold condition active.
After selecting the desired sensor calibration and pressing Enter, the next screen will ask to select
the type of temperature compensation mode desired during the calibration process. The choices
are ‘None’, ‘Standard’, ‘Light 84’, Std 75ºC’, Linear = 02.0%/ºC’ (user selectable value),
‘Glycol.5’, ‘Glycol1’, and ‘Alcohol’.
7.2 Conductivity/Resistivity Calibration
This feature provides the ability to perform a one-point or two-point Conductivity or Resistivity
‘Sensor’ calibration. The procedure described below works for both types of calibrations. There is
no reason to perform a two-point calibration on a two-electrode conductivity sensor. Four elec-
trode sensors do require a two-point calibration. It is also not practical to calibrate resistivity
sensors using (low conductivity) reference solutions. It is recommended that resistivity sensors
be sent back to the factory for calibration. Consult factory for assistance.